Home > News Center > Technical articles

Steam Sterilizer User Manual 1

                                                                                                    Steam Sterilizer User Manual  1

 Generation

Model LS-B50L Vertical Cylindrical Pressure Steam Sterilizer is controlled by microcomputer, installed with an electrical heater, a timer, an automatically pressure-temperature controller, a safety valve, a releasing valve, a pressure-temperature indicator, an alarm bell for finished sterilization and an instrument to automatically cut off the power of heater. It has the advantages of effective sterilization, 

convenient operation, safe, less electricity consuming and cheap. It is ideal equipment for clinics, scientific research institutesand other organizations to sterilizer surgical instruments, fabrics, glasses, culture media etc

Main Technical Data

CODE

NAME

LS-50LD

1

CAPACITY

50L

2

MAX.WORK PRESSURE

0.22MPa

3

MAX.WORK TEMP.

134℃

4

Heat average

≤±1℃

5

TIMER RANGE

0~99MIN

6

TEMP.RANGE

0~134℃

7

POWER / VOLTAGE

4500W /AC220V.50Hz

8

SAFETY

WATER LACK、OVER CURRENT、OVER PRESSURE

9

DIMENTION

520×520×980  (mm)

10

TRANS DIM.

590×590×1110  (mm)

11

WEIGHT

G.W.68Kg / N.W.50Kg

Structure Characteristics

1. The sterilizer component by container, lid, pipe. control system, heating system and safe systems etc.

2. The structure of the unit is a single layer cylindrical body. Its design, manufacture and inspection are strictly operated in accordance with the State Standards of Utilization Safety of Pressure Vessel (see the Quality Certificate for the details).

3. The lid is pressed once, the lock type is by screw, safe and reliable. 

4. The pipe structured by electric valve and manual water release valve, the valves reliable on movement.

5. The outline of the unit is quadrate. The displays and the control switches are centered on the controller board in the front of the unit. It is convenient.

6. The control system adopted microcomputer full automatically program processing, digital display the program.: 

1)Touch panel: sterile temperature, time, exchange temperature be adjustable and with memory function.  

2)A steam pressure stabilizer is used, the range is 0~134℃  to sterilize different objects.

3) Sterile time range 0~99hour59second be selected, when the chamber’s temperature reached the setting value, the timing auto in back way, after sterilizing, it cut off the power and alarm automatically.

4) Sterile exchange temperature range 0~110℃, for better release the cool air in chamber, and make sure the sterilizing more complete and more safety.

7. The electrical-heated elements of the unit are immersion tubes AC220v with high heating effect. It should be grounded reliably.

8. with over temperature, over pressure and lack water protection device:

1) Open type safe valve, normal it is close, when the pressure inner go over the limited, the safe valve auto open and release the steam, until the pressure back to normal it will close again for safety use.

2) With water level device, if no water or water lack occur, the alarm sound.

3) With over load protector, it will auto cut off power when short circuit, over voltage.

9. The main body is made of good quality stainless steel, and has the features of anti-corrosion, easy maintenance and long service life.